Skip to main content

X-ray beam attenuators

A compact 8-axis filter-holder fully compliant with UHV requirements

X-ray beam attenuators - IRELEC

X-ray beam attenuators

A compact 8-axis filter-holder fully compliant with UHV requirements

X-ray beam attenuators - IRELEC

The attenuator system is a compact 8-axis filter-holder fully compliant with UHV requirements.

Each filter is inserted independently into the beam path using magnetic coupling. The pneumatic actuators moving the magnets are located outside the vacuum chamber.

Two sensors indicate the actual status of each filter (in or out) and not only the pneumatic actuator positions.

IRELEC proposes this equipment with a license agreement with the European Synchrotron Radiation Facility (ESRF).

Specifications

Clear aperture

10 × 10 mm2

Overall length (flange to flange)

250 mm

Number of holders

8

No mechanical feedthrough: each filter is inserted via magnetic couplings

Independent sensors for filter in and out detection

HV or UHV compatible

KF40, CF40 or CF63 interface

References

ESRF